Cleaning The Instruments
Tools Required
• | J 23688 Coolant
and Battery Fluid Tester (Fahrenheit Scale) |
• | J 26568 Coolant
and Battery Fluid Tester (Centigrade Scale) |
Some coolant manufacturers mix other types of glycol into their coolant
formulations. Propylene glycol is the most common new ingredient.
A hydrometer does not always provide a correct measurement of freeze
protection when anything other than ethylene glycol and water is being tested.
The degree of inaccuracy varies depending on the proportion of other glycols
present in the coolant.
All new GM vehicles are produced with coolant that can be accurately
measured with a hydrometer; however, when the type and quality of coolant
being measured is unknown, such as a customer vehicle that has accumulated
mileage, the use of the refractometer is recommended.
Coolant testers J 23688
(Fahrenheit Scale) or J 26568
(Centigrade Scale) can be used to check the antifreeze protection
of the coolant. Make sure the refractometer markings are correct. Unless J 23688
or J 26568
has a provision for temperature correction,
test the temperature at which J 23688
or J 26568
is calibrated. If the coolant is warmer or cooler, the reading may be incorrect.
Follow the manufacturer's direction on using J 23688
or J 26568
.
- To clean, swing back the
plastic sample cover located at the slanted end of the J 23688
or J 26568
.
- Wipe clean both the prism and the bottom of the sample cover.
- Dry them with a tissue or a clean soft cloth.
- Close the sample cover.
Testing Procedure
- Release the clear plastic
dropper, located on the side of the J 23688
or J 26568
should be used to draw the test sample.
- Ensure that the dropper is free of any previous sample before
starting the testing procedure.
- Dip the end of the plastic dropper into the coolant without removing
it from the tester. Be sure to insert the tube slightly below the fluid level.
- Press and release the bulb to draw a sample of coolant.
- Bend the plastic tube
around the J 23688
or J 26568
so that the tip can be
inserted in the sample cover opening.
- Eject a few drops of the coolant onto the measuring prism by pressing
the bulb. Take a reading.
- The J 23688
and the J 26568
are equipped
with an eyeguard that can be used in two positions. when wearing eyeglasses,
the eyeguard should be folded back; when no eyewear is worn the eyeguard
should be extended. Proper positioning of the eyeguard will help eliminate
stray light and improve image quality.
Important: The temperature scale is reversed from a standard thermometer scale.
Readings below 0 degrees are on the upper half of the scale. The scale is
not visible until a sample is placed on the prism. If the concentration of
the solution tested is greater than the limits of the scale, the shadow will
not be visible.
- To take a reading, point
the J 23688
or the J 26568
toward any light source
and look into the eyepiece.
- The coolant freeze point
reading is taken where the dark and light portions of the scale meet. Take
the coolant reading on the right-hand scale.
- To obtain the best contrast between the light and dark portions
of the scale, tilt the J 23688
or the J 26568
toward a light source. if the edge between the light and dark shadow is not
sharp, the measuring prism was not sufficiently clean and dry or there was
not enough fluid on the measuring prism.
Calibrating Testers
The J 23688
and the J 26568
are factory calibrated
and sealed. The testers should not require adjustment. If adjustment is required,
perform the following adjustment procedure:
- Make sure the temperature of the J 23688
and the J 26568
is between 21 and 29 degrees Centigrade (70 and 85 degrees Fahrenheit).
- Using distilled water, take a reading.
- If the reading departs from a 0 degrees Centigrade (32 degrees
Fahrenheit) reading on the scale perform the following:
3.1. | Remove the sealant covering the screw. |
3.2. | Turn the screw in the direction necessary to adjust the reading
to the 0 degrees Centigrade (32 degrees Fahrenheit) line. |
3.3. | Reseal the screw using silicone sealant. |